戴安 SRS 300 离子色谱 再生抑制器
High-Capacity Electrolytically-Regenerated Suppressor for Use with IC Eluents
The SRS 300 serves virtually all analytical scale1 ion chromatography applications for both anions and cations. For anions, the preferred eluent is hydroxide. With RFIC systems, the sample is determined using conductivity detection in the lowest possible background of high-purity water. AutoSuppression means ease of use; you don’t need to make regenerant because the suppressor is constantly regenerated by the continuous electrolysis of water. The SRS 300 offers:
- Low background noise levels
- Fast startup equilibration times
- Trace anion and cation determinations
- Compatibility with mass spectrometry detection
- Compatibility with all Dionex ICS and DX chromatography modules2
- A three-fold increase in backpressure tolerance compared to previous generations
The SRS 300 delivers low backgrounds and low noise levels for the full range of Dionex ion-exchange IonPac columns and eluents, even under high-concentration gradient conditions.
When combined with the revolutionary Eluent Generator, The SRS 300 delivers truly automated, minimal-maintenance Reagent Free Ion Chromatography (RFIC).
1For capillary scale ion-chromatography applications use the Capillary Electrolytic Suppressor (CES 300).
2Chemical suppression mode only with the ICS-90 and ICS-900 systems. Older systems may require the RFC-10 or RFC-30 controller.
ordering information:
Anion Self-Regenerating Suppressor ASRS 300 4 mm | 064554 |
Anion Self-Regenerating Suppressor ASRS 300 2 mm | 064555 |
Cation Self-Regenerating Suppressor CSRS 300 4 mm | 064556 |
Cation Self-Regenerating Suppressor CSRS 300 2 mm | 064557 |
Self-Regenerating Suppressor (SRS 300) Specifications |
ASRS 300 |
Physical Specifications |
Dimensions (l × w × d) | 16.8 × 4.5 × 5.2 cm (6.6 × 1.8 × 2.1 in.) |
Weight | 630 g (1.4 lb) |
Void volume | 4 mm: <50 µL 2 mm: <15 µL |
Chemical Specifications |
Temperature Range1 | 15–35° C |
Recommended Backpressure | 30–40 psi |
Maximum Eluent Flow Rate | 3 mL/min (4 mm) 1 mL/min (2 mm) |
Eluent Solvent Restrictions2 | <40% solvents in EWM3 |
Maximum Regenerant EWM3 Flow Rate | 5 mL/min (4 mm) 2 mL/min (2 mm) |
Modes of Operation Supported | Recycle, EWM3, chemical (recycle recommended)4 |
Maximum Current | 300 mA for 4 mm, 100 mA for 2 mm |
|
CSRS 300 |
Physical Specifications |
Dimensions (l × w × d) | 16.8 × 4.5 × 5.2 cm (6.6 × 1.8 × 2.1 in.) |
Weight | 630 g (1.4 lb) |
Void volume | 4 mm: <50 µL 2 mm: <15 µL |
Chemical Specifications |
Temperature Range1 | 15–40° C |
Recommended Backpressure | 30–40 psi |
Maximum Eluent Flow Rate | 3 mL/min (4 mm) 0.75 mL/min (2 mm) |
Eluent Solvent Restrictions2 | <40% solvents in EWM3 |
Maximum Regenerant EWM3 Flow Rate | 5 mL/min (4 mm) 2 mL/min (2 mm) |
Modes of Operation Supported | Recycle, EWM3, chemical (recycle recommended)4 |
Maximum Current | 300 mA for 4 mm, 100 mA for 2 mm |
1When installed outside the heated column enclosure, all suppressors can support applications up to 60 ˚C.
2Solvents for anion eluents include methanol. Solvents for cation eluents include acetonitrile and dioxane.
3EWM = external water mode; for eluents containing >40% solvent, use the chemical regeneration mode.
4Recycle recommended for aqueous applications without solvent.
The SRS 300 is compatible with Dionex IC systems equipped with a suppressor power supply. The new SRS 300 products directly replace the SRS ULTRA II. Simply install the new SRS 300 into your current Dionex IC system.
Ordering Information |
Anion Self-Regenerating Suppressor ASRS 300 4 mm | 064554 |
Anion Self-Regenerating Suppressor ASRS 300 2 mm | 064555 |
Cation Self-Regenerating Suppressor CSRS 300 4 mm | 064556 |
Cation Self-Regenerating Suppressor CSRS 300 2 mm | 064557 |