赛默飞Thermo RFIC-Eluent Generation离子色谱IC柱
品牌 | 厂商性质 | 产地 | 货期 |
---|---|---|---|
赛默飞 | 一般经销商 | 美国 | 现货 |
性能特点
EGC cartridges support a concentration range of 0.1–100 mM (0.1–80 mM EGC-LiOH) at 1.0 mL/min. T
Eluent Generation—A Major Component of Reagent-Free IC
Eluent generation prevents baseline shift, increases sensitivity, improves resolution, and ensures consistent peak integration. A range of Thermo Scientific™ Dionex™ EGC cartridges are available for the production of hydroxide, carbonate, bicarbonate and methanesulfonic acid eluents. Eluent generation eliminates the need to handle acids and bases traditionally required for the preparation of IC eluents, and allows chromatographers to run a full range of gradient and isocratic separations more effectively than hand-made eluents.
Automatic production of high-purity eluents for isocratic and gradient separations on Thermo Scientific Dionex ICS-5000+ and ICS-2100 systems
Outstanding run-to-run reproducibility week after week
Elimination of acid and base handling
Longer lasting pumps because they encounter only deionized water
Supports both Analytical and Capillary flow rates
Supports HPIC™ system pressures up to 5000 psi (34.5 MPa, Dionex EGC Capillary and EGC 500 only)
Dionex EGC III and EGC 500 cartridges support a concentration range of 0.1–100 mM (0.1–80 mM EGC III LiOH, 0.1-15 mM EGC 500 K2CO3) at 1.0 mL/min. The KOH, NaOH and LiOH cartridges support a maximum of 25% methanol. The K2CO3 and MSA cartridges are not solvent-compatible.
Dionex EGC (Capillary) cartridges support a concentration range of 0.1–200 mM at 0.010 mL/min and a flow rate range of 0.001–0.030 mL/min. The Dionex EGC-KOH (Capillary) cartridge supports a maximum of 25% methanol. The Dionex EGC-MSA(Capillary) cartridge is not solvent-compatible.
RFIC-EG Component Specifications | |
---|---|
EGC 500 Cartridge Specifications | |
Cartridge Dimensions (h × diam.) | 25.4 × 10.8 cm (10 × 4.25 in.) |
Cartridge Weight | 1.4 kg (3.0 lb) |
Concentration Range | Dionex EGC 500 KOH: 0.1–100 mM Dionex EGC 500 MSA: 0.1–100 mM Dionex EGC 500 K2CO3: 0.1–15 mM |
Flow Rate | 0.10–3.00 mL/min. |
Max. Operating Pressure | 34.5 MPa (5000 psi) |
Max. Solvent Concentration | Dionex EGC 500 KOH: 25% methanol Dionex EGC 500 MSA: no solvents Dionex EGC 500 K2CO3: no solvents |
EGC III Cartridge Specifications | |
Cartridge Dimensions (h × w × d): | 23 × 7 × 10 cm (9 × 2.75 × 4 in.) |
Cartridge Weight: | 1.6 kg (3.5 lb) |
Concentration Range: | Dionex EGC III KOH, NaOH: 0.1–100 mM Dionex EGC III LiOH: 0.1–80 mM |
Flow Rate: | 0.01–3.00 mL/min. |
Max. Operating Pressure: | 20.7 MPa (3000 psi) |
Max. Solvent Concentration: | Dionex EGC III KOH, NaOH, LiOH —25% methanol Dionex EGC III MSA — no solvents |
EGC (Capillary) Cartridge Specifications | |
Cartridge Dimensions (h × diam.): | 15 × 6.4 cm (6 × 2.5 in.) |
Cartridge Weight: | 0.375 kg (0.83 lb.) |
Concentration Range: | 0.1-200 mM |
Flow Rate: | 1-30 µL/min (0.001-0.030 mL/min) |
Max. Operating Pressure: | 34.5 MPa (5,000 psi) |
Max. Solvent Composition: | EGC-KOH - 25% MeOH EGC-MSA - no solvents |
RFIC CR-ATC and CR-CTC Specifications | |
Dimensions (h × w × l): | 5.1 cm × 5.5 cm × 8.4 cm (2.0 in × 2.15 in × 3.3 in) |
Weight: | 60 g (0.13 lb) |
Current Output (Analytical): | <125 mA |
Void Volume (Analytical): | <100 μL |
Constant Voltage (Analytical): | 24 V dc |
Current Output (Capillary): | 1 mA |
Void Volume (Capillary): | < 2 μL |
Constant Voltage (Capillary): | 12 V dc |
RFIC Suppressor Specifications | |
AERS 500 and CERS 500 | |
Dimensions (h × w × l): | 12.1 × 4.5 × 4.8 cm (4.25 × 1.8 × 1.9 in.) |
Void Volume: | 4 mm: <50 μL 2 mm: <15 μL |
Weight: | 295 g (0.65 lb) |
Current Range: | 0–500 mA |
AAES and CAES | |
Dimensions (h × w × l): | 4.9 × 4.4 × 10.2 cm (1.9 × 1.8 × 4.0 in.) |
Void Volume: | <35 μL |
Weight: | 120 g (0.3 lb) |
ACES 300 and CCES 300 | |
Dimensions (h × w × l): | 9.4 × 3.1 × 10.3 cm (3.7 × 1.2 × 4.1 in.) |
Void Volume: | < 1.5 μL |
Weight: | 150 g (0.33 lb.) |
Current Range: | 0-20 mA |
RFC -30 Reageant-Free Controller Specifications | |
---|---|
TTL Inputs: | Two independent 0–5 V 0V = On, 5V = Off; ac control |
Dimensions (h × w × d): | 12.4 × 16.2 × 28.8 cm (4.9 × 6.4 × 11.3 in.) |
Weight: | 2.5 kg (5.5 lb) |
Power Requirements: | Consumption, 500 VA max.; voltage, 100–240 V ac; frequency, 50/60 Hz |
Operating Temperature Range: | 4–40 °C |
Operating Humidity Range: | 5–95% relative, noncondensing |
Accessories | |
---|---|
Potassium Hydroxide Eluent Generator Cartridge (EGC III KOH) | 074532 |
Sodium Hydroxide Eluent Generator Cartridge (EGC III NaOH) | 074533 |
Lithium Hydroxide Eluent Generator Cartridge (EGC III LiOH) | 074534 |
Sodium Carbonate Eluent Generator Cartridge (EGC 500 K2CO3) | 088453 |
Dionex Eluent pH Modifier (EPM 500) | 088471 |
Dionex EGC 500 K2CO3 Carbonate Mixer Column (2 mm) | 088467 |
Dionex EGC 500 K2CO3 Carbonate Mixer Column (4 mm) | 088468 |
Methanesulfonic Acid Eluent Generator Cartridge (EGC III MSA) | 074535 |
EGC-KOH (Capillary) Cartridge | 072076 |
EGC-MSA (Capillary) Cartridge | 072077 |
Potassium Hydroxide Eluent Generator Cartridge (EGC 500 KOH) | 075778 |
Methanesulfonic Acid Eluent Generator Cartridge (EGC 500 MSA) | 075779 |
赛默飞Thermo RFIC-Eluent Generation离子色谱IC柱信息由上海希言科学仪器有限公司为您提供,如您想了解更多关于赛默飞Thermo RFIC-Eluent Generation离子色谱IC柱报价、型号、参数等信息,欢迎来电或留言咨询。
注:该产品未在中华人民共和国食品药品监督管理部门申请医疗器械注册和备案,不可用于临床诊断或治疗等相关用途